Model Number 5392 |
Device Problems
Mechanical Problem (1384); Battery Problem (2885); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: analysis could not confirm the customer comment of the external pulse generator powering itself off and on by itself.The device was contaminated in the output connector assembly, main printed circuit board (pcb), upper- and lower-case halves, keypad flex, encoder assembly flex, battery drawer, battery drawer shorting bar and liquid crystal display (lcd) assembly flex.The hanger was broken.The errors found in log; main pcb is out of specification (electrical).Both wires on the lcd were pinched (not compromised).The batteries received with device were both at 1.45 volts no load.All found defective parts were replaced and all other identified issues were resolved.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator needed repair.It was further reported that during use the battery indicator would randomly turn red and the device would power cycle.It was noted that new batteries did not resolve the issue.The device has been returned for service.No patient involvement in the event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|