• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM Back to Search Results
Catalog Number 256082
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
Eua #: (b)(4).Initial reporter address 1: customer has declined to provide facility address, or facility name, however, it was located in (b)(6).Available corporate info is being used.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported while testing for sars-cov-2 a false positive result was obtained.Repeat testing performed using pcr test method, and the result was negative.There was no report of patient impact.Eua #: (b)(4).
 
Manufacturer Narrative
Eua#: (b)(4).H6: investigation summary bd has received several customer complaints for false positive results, when using bd sars-cov-2 reagents for bd veritor¿ system.The current investigation concerns multiple lots of bd sars-cov-2 reagents for bd veritor¿ system.Bd takes a systematic approach to investigating false positive complaints that are received.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples if applicable.Retain samples were tested for batch #: 0224483.The defect could not be replicated and therefore the complaint could not be confirmed.A device history review was conducted for lot number 0224483.The investigation did not find a root cause for the false positive results that were observed.It is recommended that each customer review their workflow carefully to ensure that the package insert is being followed as written.The root cause is under investigation and will be documented in our quality system.Bd cannot confirm the complaint based on the investigation that was performed.A corrective and preventive action (capa#1878253) is already initiated to investigate the root cause and some mitigation actions are already being addressed.H3 other text : see h10.
 
Event Description
It was reported while testing for sars-cov-2 a false positive result was obtained.Repeat testing performed using pcr test method and the result was negative.There was no report of patient impact.Eua # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RAPID DETECTION OF SARS-COV-2 VERITOR
Type of Device
CORONAVIRUS ANTIGEN DETECTION SYSTEM
Manufacturer (Section D)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH 
MDR Report Key10960205
MDR Text Key223462979
Report Number3006948883-2020-00903
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/27/2021
Device Catalogue Number256082
Device Lot Number0224483
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/18/2020
Initial Date FDA Received12/07/2020
Supplement Dates Manufacturer Received07/29/2021
Supplement Dates FDA Received08/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-