Model Number URF-V2
Device Problems Leak/Splash (1354); Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device was received and evaluated.Evaluation determined that the device was found with cut on the ¿a-rubber¿ and a loose bcu (body control unit) was observed.Broken bending section was determined and intermittent image was observed.In addition, peeling on ¿a-rubber¿ glue was determined and switches are found not working.The components of the device were repaied and restored to olympus original factory specifications.Based on evaluation findings the reported failure was confirmed and failures found likely attributed to users handling and or maintenance issue.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
It was reported the device failed insulation test.The issue occurred during reprocessing.There was no patient involvement/impact reported.No user injury reported.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10961042
MDR Text Key226928678
Report Number8010047-2020-09956
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/11/2020
Initial Date FDA Received12/07/2020
Supplement Dates Manufacturer Received01/14/2021
Supplement Dates FDA Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-