• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problems Energy Output Problem (1431); Failure to Read Input Signal (1581); Application Program Problem (2880)
Patient Problem Burn, Thermal (2530)
Event Date 10/08/2020
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during a cataract procedure, a patient experienced a small burn in the main incision.Two sutures were required to close the wound.A contact lens was also placed on the eye.The customer believes there is a potential mismatch between the programmed ultrasound power delivered causing the event.
 
Manufacturer Narrative
Additional information provided in h.6.And h.10.A review of the service records (sr) showed that there was no sr open for this event.It should be noted that on (b)(6) 2020, an sr was opened for preventative maintenance (pm) was performed.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.With no additional, related information provided, the customer reported events could not be confirmed.The root cause of the reported events cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10968080
MDR Text Key220205573
Report Number2028159-2020-01070
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number8065750833
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/16/2020
Initial Date FDA Received12/08/2020
Supplement Dates Manufacturer Received12/10/2020
Supplement Dates FDA Received12/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-