NIHON KOHDEN CORPORATION ORG-9110A; MULTIPLE PATIENT RECEIVER Back to Search Results
Model Number ORG-9110A
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2020
Event Type  malfunction  
Manufacturer Narrative
The customer reported that their multiple patient receiver (org) lost communication with all monitored transmitters.The customer rebooted the org, but the issue persisted.The customer then rebooted the org switch, and all the devices came back up.No harm or injury was reported.Nihon kohden continues to investigate the reported event.Concomitant medical device: the following device was used in conjunction with the org: multiple transmitters: model: ni.Sn: ni.
 
Event Description
The customer reported that their multiple patient receiver (org) lost communication with all monitored transmitters.
 
Manufacturer Narrative
Details of complaint: the customer reported that the multiple patient receiver (org) lost communication with all monitored transmitters.The customer rebooted the org, but the issue persisted.The customer then rebooted the org switch, and all the devices came back up.No patient harm was reported.Service requested / performed: troubleshooting.Investigation summary: nihon kohden tech support (nk ts) was able to identify that the communication loss was occurring due to the customer's network.Nk ts was able to walk the customer through several troubleshooting steps to resolve the issue.As the issue of communication loss was resolved without any rework of nk devices, the cause is unlikely to be related to any malfunction.It was identified that the customer also had non-nk devices connected to the nk network which was causing network congestion.The root cause for communication loss is most likely related to the customer's network.
 
Event Description
The customer reported that the multiple patient receiver (org) lost communication with all monitored transmitters.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORG-9110A
Type of Device
MULTIPLE PATIENT RECEIVER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key10969224
MDR Text Key238986017
Report Number8030229-2020-00739
Device Sequence Number1
Product Code DRG
UDI-Device Identifier04931921103906
UDI-Public04931921103906
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORG-9110A
Device Catalogue NumberORG-9110A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2020
Initial Date FDA Received12/08/2020
Supplement Dates Manufacturer Received06/03/2022
Supplement Dates FDA Received11/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MULTIPLE TRANSMITTERS; MULTIPLE TRANSMITTERS
-
-