• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY ORBERA INTRAGASTRIC BALLOON SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY ORBERA INTRAGASTRIC BALLOON SYSTEM Back to Search Results
Model Number B-50000
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Abdominal Pain (1685); Vomiting (2144)
Event Date 10/10/2020
Event Type  malfunction  
Manufacturer Narrative
A review of the device labeling notes the following: the risk documentation for this device establishes the occurrence ranking for the reported event "inflation" as "remote", which is defined by apollo as.02% -.49% of complaints over units sold.A review for the intragastric balloon products for the failure mode "inflation" is performed during quarterly complaint analysis meetings (cam) and has demonstrated that global balloon inflation rate remains at "remote".Therefore, apollo determined that the reported event is occurring within the range of the expected frequency and severity for this reported event, as referenced in the cam meeting slides.The current orbera® intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of "pain", "vomiting" as follows: "the physiological response of the patient to the presence of the orbera system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response." "each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms." "complications: possible complications of the use of the orbera system include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach or as a result of the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus." orbera® system placement and inflation (step-by-step) - slowly fill the balloon with sterile saline, 50 cc at a time.Repeat up to 700 cc (14 strokes).Recommended full volume is up to 700 cc.Minimum fill volume is 400 cc.
 
Event Description
Patient presented with pain and vomiting.The balloon hyperinflated.
 
Manufacturer Narrative
Supplement #1 submitted to the fda on 03/feb/2021.Device evaluation summary: the device was returned to the apollo device analysis laboratory on 12/jan/2021.A deflated balloon with a significant amount of blue/green discoloration and fungus on the shell was returned.There are several slits on the shell observed with the naked eye.Under microscopic analysis, the openings on the shell were noted to have striated edges, consistent with damage from a surgical tool for removal purposes.Functional evaluation could not be conducted due to the slits on the shell.The complaint could not be verified as it is uncertain how the inflation occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key10969236
MDR Text Key221288491
Report Number3006722112-2020-00121
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020541
UDI-Public(01)10811955020541(17)2021-07-02(10)AF03046
Combination Product (y/n)N
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/02/2021
Device Model NumberB-50000
Device Catalogue NumberB-50000
Device Lot NumberAF03046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Initial Date Manufacturer Received 11/12/2020
Initial Date FDA Received12/08/2020
Supplement Dates Manufacturer Received01/12/2021
Supplement Dates FDA Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient Weight87
-
-