• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number H93811
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 11/10/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the spikes of three (3) sterile repeater pump tube sets separated from the tubing, leading to a leak.This occurred during compounding.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Correction to the previously submitted g4: date received by mfr.G4: the correct baxter aware date is 12/07/2020, previously submitted as 11/10/2020.H4: the lot was manufactured between july16, 2020 - july 17,2020.H10: two actual devices were received for evaluation.Visual inspection was performed and the spike was found detached from the damaged tube set tubing of both samples.Functional testing was not performed.The reported condition was verified for both samples.The cause of the condition was not determined, however the most likely cause was due to inadequate or lack of loctite 4601 adhesive being applied during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The remaining device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, STER
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10970828
MDR Text Key220689921
Report Number1416980-2020-07641
Device Sequence Number1
Product Code NEP
UDI-Device Identifier00085412475721
UDI-Public(01)00085412475721
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2023
Device Catalogue NumberH93811
Device Lot Number60249189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2020
Initial Date Manufacturer Received 11/10/2020
Initial Date FDA Received12/08/2020
Supplement Dates Manufacturer Received12/07/2020
Supplement Dates FDA Received12/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-