• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. FA QUANTUM 2 CONTROLLER; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. FA QUANTUM 2 CONTROLLER; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 28168
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2020
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case- (b)(4).The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection observed the footpedal port is damaged preventing full plugin.A functional evaluation revealed that most settings could not be tested due to the damaged footpedal port.The complaint was verified.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include misalignment of the connector when connecting to the unit receptacle in which excessive force or twisting could cause damage.No containment or corrective actions are recommended at this time. there are no indications to suggest that the device/ product did not meet specifications upon release into distribution.
 
Event Description
It was reported that during procedure, the controller was not working.No delay and a back up was available to complete the procedure.No other complications were reported.No further information.Results of investigation revealed that during functional evaluation the controller could not be tested due to the damaged footpedal port which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FA QUANTUM 2 CONTROLLER
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10973186
MDR Text Key220901704
Report Number3006524618-2020-01076
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082666
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number28168
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-