Model Number 102096-67A |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
|
Event Description
|
On (b)(6) 2020,senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
|
|
Manufacturer Narrative
|
The sensor triggered a true led disconnect alarm as it was determined the sensor led would not light up and the led during the qc test experienced an intermittent crash.Investigation showed that the system disabled the sensor due to a detected performance failure and the system's self-test functions were still working normally when the issue happened.Device available for evaluation? yes, device received on 10 december 2020 device evaluated by manufacturer? yes (b)(4).
|
|