Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ft4 iii (ft4 iii) on a cobas e801 module compared to the abbott architect method.The results from the e801 module were reported outside of the laboratory where additional testing was requested.The sample was submitted for investigation where discrepant ft4 iii results were also identified by the centaur method and an e801 module used at the investigation site.Refer to attached data for the patient results.The customer¿s e801 module serial number was (b)(4).The e801 module used at the investigation site was 14d9-06.The ft4 iii reagent lot number used at the investigation site was 460793 with an expiration date of feb-2021.
 
Manufacturer Narrative
Calibration and qc at the investigation site were acceptable.The sample was submitted for investigation.No interfering factor was identified.Assays from different manufacturers can generate different results.This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10979236
MDR Text Key220653881
Report Number1823260-2020-03157
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/16/2020
Initial Date FDA Received12/09/2020
Supplement Dates Manufacturer Received11/16/2020
Supplement Dates FDA Received03/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-