ARTHROCARE CORP. FIRSTPASS MINI RIGHT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72290130
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case- (b)(4).
 
Event Description
It was reported that during a knee meniscus root repair, the first pass mini tape passed successfully the first time through the meniscus but the second tape failed to capture suture 5 times.The procedure was completed with a backup device and no delay or further complications reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the reported device used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned device shows suture trap is broken.A functional evaluation shows the needle will deploy when trigger is initiated.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRSTPASS MINI RIGHT
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10979654
MDR Text Key220653187
Report Number3006524618-2020-01082
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00885556694572
UDI-Public00885556694572
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/11/2022
Device Model Number72290130
Device Catalogue Number72290130
Device Lot Number2030309
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2021
Initial Date Manufacturer Received 11/20/2020
Initial Date FDA Received12/09/2020
Supplement Dates Manufacturer Received03/31/2021
Supplement Dates FDA Received04/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-