• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problems Use of Device Problem (1670); Device Dislodged or Dislocated (2923)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 11/25/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient was revised due to loosening of the cup and liner at the bone to implant interface.Unknown cement was used.It was also reported that the patient's liner was worn and loose from cup, not sure if it disassociated first or wore out then came loose.Cup is vertical, and patient is morbidly obese.That would seem to contribute to this issue so surgeon removed cup repositioned a new cup.Doi:1st revision: (b)(6) 2012; doi: 2nd revision: (b)(6) 2018; dor: (b)(6) 2020; left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a device history record (dhr) review, was not possible because the required lot code(s) was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR CUP
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10983401
MDR Text Key220669825
Report Number1818910-2020-26691
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/25/2020
Initial Date FDA Received12/10/2020
Supplement Dates Manufacturer Received01/12/2021
Supplement Dates FDA Received01/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALTRX +4 10D 32IDX50OD; ARTICUL/EZE BALL 32 +1 GR; UNKNOWN HIP ACETABULAR CUP; ALTRX +4 10D 32IDX50OD; ARTICUL/EZE BALL 32 +1 GR; UNKNOWN HIP ACETABULAR CUP
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient Weight124
-
-