THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
|
Back to Search Results |
|
Model Number 201-10002 |
Device Problems
Mechanical Problem (1384); Noise, Audible (3273)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
|
|
Event Description
|
It was reported that the centrimag motor started making more noise than normal (a "mechanical noise") during the preparation for transporting an ecmo patient.A low flow alarm appeared on the console.The perfusionist immediately replaced the motor drive and the issue was resolved.The motor drive was to be returned.
|
|
Manufacturer Narrative
|
Manufacturer's investigation conclusion: the reported event of motor noises was not confirmed.The centrimag motor was returned for analysis to the european distribution center (edc) and was functionally tested at different speed for a period of 72 hours.No anomalies were found, and the motor was found to function as intended.The reported event was unable to be reproduced during testing.Preventative maintenance was performed before being returned to the customer.The root cause for the reported event was unable to be conclusively determined through this analysis.The device history records were reviewed for the centrimag motor and the motor was found to pass all manufacturing and qa specifications before being shipped to the customer on 27nov2019.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction.".The 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual section 12.1 entitled "appendix i ¿ primary console alarms and alerts" contains a list of console alarms and alerts, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing this event.
|
|
Search Alerts/Recalls
|
|
|