Model Number RMU-1000 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
Death (1802)
|
Event Date 11/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Although requested, the device associated with this event has not been returned.The investigation is on-going and the cause for the event is not known.Should additional information become available a follow-up mdr shall be submitted.
|
|
Event Description
|
It was reported by an ems department that when they arrived on scene and powered on the compression module, its service and warning lights were on and it would not perform compressions.They turned it off then back on with the same result.Cpr was performed manually, and the patient never responded.No additional event or patient information was provided.Although there was no connection reported between the device malfunction and the outcome of the patient, as manual compressions were performed, this mdr is being submitted out of an abundance of caution.
|
|
Manufacturer Narrative
|
Analysis of the returned device identified the root cause as being related to an open trace on the arm power motor drive pcb which caused the battery (rbp-g100gg) to drain approximately twice as fast.
|
|
Manufacturer Narrative
|
Analysis of the returned device identified the root cause as being related to a component failure on the arm power board pcb, specifically the power controller chip.
|
|