ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/18/2020
Event Type  malfunction  
Manufacturer Narrative
The calibration and qc data was requested but not provided.The investigation did not identify a product problem.The cause of the event could not be determined.Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and differences in the standardization methodology used.This event occurred in (b)(6).
 
Event Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ft3 iii (ft3 iii) and elecsys ft4 iii (ft4 iii) on a cobas e801 module compared to the abbott architect method.The results from the e801 module were reported outside of the laboratory where additional testing was requested.The sample was submitted for investigation where discrepant ft4 iii results were also identified by the centaur method and an e801 module used at the investigation site.This medwatch will cover ft4 iii.Refer to medwatch with patient identifier (b)(4) for information on the ft3 iii results.Refer to the attached data for the patient results.The ft4 iii reagent lot number used at the investigation site was 460793 with an expiration date of feb-2021.The customer¿s e801 module serial number was (b)(4).The e801 module serial number used at the investigation site was (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10992124
MDR Text Key220972871
Report Number1823260-2020-03177
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2020
Initial Date FDA Received12/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
THYRADIN
-
-