• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH REDUC-FORC W/POINT SOFTLO L155; FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH REDUC-FORC W/POINT SOFTLO L155; FORCEPS Back to Search Results
Catalog Number 399.086
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Event date is unknown.Udi number is unknown.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6).Reports an event as follows: it was reported on an unknown date that during an unknown procedure, the reduction forceps tip got broken.It was unknown if the surgery completed successfully.The patient outcome was unknown.This compliant involves one (1) device.This report is for (1) reduc-forc w/point softlo l155 this report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Investigation summary: based on the received pictures, complaint is confirmed as it is visible that one prong at the tip is broken of as described in complaint description.Otherwise it seems that the instrument is in a good condition.Product was not returned therefore no further investigation possible.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available (information or/and material), the investigation will be updated as applicable.Device history lot: part: 399.086.Lot: 5941589.Manufacturing site: salzburg.Release to warehouse date: 08.May.2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d4: udi number g4: device is not distributed in the united states, but is similar to device marketed in the usa.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.G1: corrected physical manufacturer.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: the overall complaint was confirmed for the received device.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part: 399.086.Lot: 5941589.Manufacturing site: salzburg.Release to warehouse date: 08.May.2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUC-FORC W/POINT SOFTLO L155
Type of Device
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SALZBURG
vorgartenstrasse 206b
salzburg 1020 WIEN
AU   1020 WIEN
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key10994697
MDR Text Key246625176
Report Number8030965-2020-09637
Device Sequence Number1
Product Code HTD
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number399.086
Device Lot Number5941589
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received12/12/2020
Supplement Dates Manufacturer Received12/16/2020
01/29/2021
12/08/2021
01/07/2022
Supplement Dates FDA Received01/11/2021
02/03/2021
01/04/2022
02/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-