The patient was undergoing a thrombectomy procedure in the right tibial artery using an indigo system catrx aspiration catheter (catrx), aspiration tubing (tubing), and a non-penumbra sheath.During the procedure, while advancing the catrx through the sheath, the physician initially experienced resistance but was able to track the catrx down to the posterior tibial artery.However, when aspiration was turned on, the physician noticed that there was no flow through the tubing.Subsequently, while removing the catrx, the physician experienced resistance and the proximal end of the catrx was found to be kinked.The procedure was completed using a new catrx and the same sheath.There was no report of an adverse effect to the patient.
|
Evaluation of the returned catrx revealed a fracture.This damage was likely the reported kink.If the device is forcefully manipulated against resistance, damage such as a kink and subsequent fracture may occur.Further evaluation revealed an ovalization on the distal shaft.If the device is forcefully pinched or gripped or is otherwise mishandled during use, damage such as an ovalization may occur.This damage may have contributed to resistance during the procedure.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|