Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the instrument fractured during a procedure.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information is available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Three of item 09-0257 lot b20 were returned and verified per etch.Visual inspection of returned items show the tips of the instruments are fractured in varying degrees.Visual of the broken tips did not show any other signs that are significant.Scratching and staining of instruments from previous uses was observed.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|