• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC KETONE TEST STRIPS; NITROPRUSSIDE, KETONES (URINARY, NON-QUANT.)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC KETONE TEST STRIPS; NITROPRUSSIDE, KETONES (URINARY, NON-QUANT.) Back to Search Results
Model Number STRIP, WALGREENSKETONE 50CT #496636
Device Problem Material Discolored (1170)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
Internal report reference number: (b)(4).Ketone test strips were returned: defect was detected.Root cause: rc-067: storage outside specifications.Note: manufacturer contacted customer in several follow-up calls to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
 
Event Description
Consumer reported complaint for physical defect of ketone test strips, stating when she had opened the vial, the padding on the test strips was gray.Customer is using the product for keto diet purposes.Customer stated that when she had attempted to use the product, the padding had remained gray.The package had not been open or damaged when received and there was no evidence of the product being previously used.The customer feels well and did not report any symptoms.No medical attention was reported.Customer declined to perform a test at the time of the call.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KETONE TEST STRIPS
Type of Device
NITROPRUSSIDE, KETONES (URINARY, NON-QUANT.)
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key10999145
MDR Text Key229565633
Report Number1000113657-2020-00895
Device Sequence Number1
Product Code JIN
UDI-Device Identifier021292008178
UDI-Public(01)021292008178
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 12/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2021
Device Model NumberSTRIP, WALGREENSKETONE 50CT #496636
Device Lot NumberAW507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Distributor Facility Aware Date11/17/2020
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received12/13/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-