MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION SILVER SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION SILVER SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753008
Device Problem Component Missing (2306)
Patient Problem Insufficient Information (4580)
Event Date 11/01/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported no ozil with the ophthalmic system.Procedure details and patient impact where not reported.
 
Manufacturer Narrative
The company service representative examined the system and was not able to confirm or replicate the reported event.The system software was upgraded as a preventative measure (pm).The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The system was found to meet specifications; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SILVER SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11002437
MDR Text Key221802904
Report Number2028159-2020-01108
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753008
Device Lot Number13D186
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/19/2020
Initial Date FDA Received12/14/2020
Supplement Dates Manufacturer Received02/18/2021
Supplement Dates FDA Received03/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-