BOSTON SCIENTIFIC CORPORATION PERCUFLEX NEPHROURETERAL STENT; STENT, URETERAL Back to Search Results
Model Number 35095
Device Problems Material Deformation (2976); Material Split, Cut or Torn (4008)
Patient Problems No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2020
Event Type  Injury  
Manufacturer Narrative
Date of event was approximated to (b)(6) 2020 as the event date occured two days after being placed inside the patient.The complainant was unable to report the device upn and lot number; therefore, the manufacture date and expiration date are unknown.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
Note: this report pertains to one of four percuflex locking loop nephrostomy catheters used in the same patient.It was reported to boston scientific corporation that two percuflex locking loop nephrostomy catheters were place into the patient during a procedure performed on (b)(6) 2020.According to the complainant, on (b)(6) 2020, it was noted that both of the catheters were cracked and leaking.Reportedly, the patient had to be transported through medivac for urgent replacement procedure of the catheters two weeks post insertion.Few days after the leakage was noted, a replacement procedure was completed with another two percuflex locking loop nephrostomy catheters; however, the same issue occured with the newly implanted catheters within 2 days inside the patient.Patient is well but scheduled for another tube replacement week of (b)(6).There were no patient complications reported as a result of these events.
 
Event Description
Note: this report pertains to one of four percuflex locking loop nephrostomy catheters used in the same patient.It was reported to boston scientific corporation that two percuflex locking loop nephrostomy catheters were place into the patient during a procedure performed on (b)(6), 2020.According to the complainant, on (b)(6), 2020, it was noted that both of the catheters were cracked and leaking.Reportedly, the patient had to be transported through medivac for urgent replacement procedure of the catheters two weeks post insertion.Few days after the leakeage was noted, a replacment procedure was completed with another two percuflex locking loop nephrostomy catheters; however, the same issue occured with the newly implanted catheters within 2 days inside the patient.Patient is well but scheduled for another tube replacement week of december 15 to 18.There were no patient complications reported as a result of these events.**additional information received on (b)(6) 2020** reportedly, the devices were used from kidney, through the ureter and into the bladder.
 
Manufacturer Narrative
Block b3: date of event was approximated to (b)(6) 2020 as the event date occured two days after being placed inside the patient.Block d4, h4: the complainant was unable to report the device upn and lot number; therefore, the manufacture date and expiration date are unknown.Block h6: patient code 3191 captures the reportable event of additional procedure required to replace the defective device.Block h10: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.Additional information: block b5.
 
Manufacturer Narrative
Block b3: date of event was approximated to (b)(6) 2020 as the event date occurred two days after being placed inside the patient.Device evaluated by manufacturer: the device was not returned for analysis.Photos from the procedure were provided and reviewed.The images showed the device was implanted in the patient and a shaft kink was visible next to the patient's body.
 
Event Description
Note: this report pertains to one of four percuflex locking loop nephrostomy catheters used in the same patient.It was reported to boston scientific corporation that two percuflex locking loop nephrostomy catheters were place into the patient during a procedure performed on (b)(6) 2020.According to the complainant, on (b)(6) 2020, it was noted that both of the catheters were cracked and leaking.Reportedly, the patient had to be transported through medivac for urgent replacement procedure of the catheters two weeks post insertion.Few days after the leakeage was noted, a replacment procedure was completed with another two percuflex locking loop nephrostomy catheters; however, the same issue occured with the newly implanted catheters within 2 days inside the patient.Patient is well but scheduled for another tube replacement week of (b)(6).There were no patient complications reported as a result of these events.Reportedly, the devices were used from kidney, through the ureter and into the bladder.It was further reported the replacement procedure occurred on (b)(6) 2020 and non-boston scientific catheters were used.The four boston scientific catheters used prior were percuflex nephroureteral stents.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUFLEX NEPHROURETERAL STENT
Type of Device
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11002488
MDR Text Key221255092
Report Number3005099803-2020-06078
Device Sequence Number1
Product Code FAD
Combination Product (y/n)N
PMA/PMN Number
K924608
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2023
Device Model Number35095
Device Catalogue Number35095
Device Lot Number0025991829
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/18/2020
Initial Date FDA Received12/14/2020
Supplement Dates Manufacturer Received12/22/2020
01/18/2021
Supplement Dates FDA Received01/12/2021
02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-