• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA PLATELET+SAMPLER, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA PLATELET+SAMPLER, PLASMA, RBC SET Back to Search Results
Model Number 82446
Device Problems Partial Blockage (1065); High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6).The platelet collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Root cause :based on the findings from the run data file analysis, the leukodreduction failure was caused by an air block in the plasma line.The airblock likely occurred when the centrifuge was stopped during blood prime.The excess plasma flow destabilized the cell bed in the lrs chamber causing some wbcs to escape from the chamber.
 
Manufacturer Narrative
Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The run data file (rdf) was analyzed for this event.Analysis of the run data file determined there was an airblock in the plasma line for the entire collection, likely introduced when the centrifuge was stopped during blood prime.Having the plasma line blocked caused more fluid to be drawn through the lrs chamber since we were missing the plasma pump contribution.And since the procedure looked the same for the whole collection, there wasn't a change in signal to the rbc detector to induce the flag.Investigation is in process.A follow up report will be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA PLATELET+SAMPLER, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key11003426
MDR Text Key222289914
Report Number1722028-2020-00546
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583824461
UDI-Public05020583824461
Combination Product (y/n)N
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2022
Device Model Number82446
Device Catalogue Number82446
Device Lot Number2008034151
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 11/20/2020
Initial Date FDA Received12/14/2020
Supplement Dates Manufacturer Received01/29/2021
02/23/2021
Supplement Dates FDA Received02/18/2021
03/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-