DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD
Device Problems Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Necrosis (1971); Pain (1994); Osteolysis (2377); Ambulation Difficulties (2544)
Event Date 11/13/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Revision of the right hip implant after pain and difficulty in walking.During the revision it was detected abundant reaction tissue with joint effusion blood serum from suspected serum/pseudotumor reaction.The histological examination showed an intense inflammatory reaction from a foreign body to a predominant macrophage component of the bone and periosteal soft tissues, conditioning aseptic osteolysis.Doi: (b)(6) 2007; dor: (b)(6) 2019; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
The prosthesis implanted by the patient has released metal particles in the implant site.It also stated that,necrotic tissue reactive from wear of the prosthetic components were sent for histological examination.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11004473
MDR Text Key221501843
Report Number1818910-2020-26889
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,other
Type of Report Initial,Followup,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/14/2020
Supplement Dates Manufacturer Received12/02/2020
12/23/2020
Supplement Dates FDA Received12/18/2020
01/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL STEM; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD
Patient Outcome(s) Required Intervention;
-
-