Device was received and evaluated.Evaluation of the device was found with cloudy image due to damage on distal tip leading to breakage of image bundle.In addition, the lg (light guide) fibers were observed to be broken.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Based on device evaluation results, the damages found on the device were most likely due to user mishandling.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.Olympus will continue to monitor complaints for this device.
|