This supplemental report is being submitted to provide review of the device history records (dhr),device evaluation and investigation conclusion.Please see updated sections: d4, d10,g4, g7, h2, h3,h4, h6 and h10.Device evaluation of the returned device, the reported issue was confirmed.It was found that the unit was leaking from block body and tube assay, causing the actuator block to spark.Burnt actuator block was observed.Dhrs for this product have been reviewed.All records indicate that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The root cause of the reported failure cannot be determined at this time.Per instruction for use: the use of cords other than those specified may result in increased emissions or decreased immunity of the medical electrical equipment.Use only surgical gloves specifically designated by their manufacturer for electrosurgical procedure.In addition, examine the device prior to use.Do not use if damage is found.Olympus will continue to monitor complaints for this device.
|