• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO ADVANTA 2 FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO ADVANTA 2 FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P1190A000067
Device Problem Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Date 12/09/2020
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the siderail interface board, motor control board and left siderail cable needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in july 2020.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the siderail interface board, motor control board and left siderail cable to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the head section was moving on its own.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
harish vishwanathan
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key11012474
MDR Text Key223183272
Report Number3006697241-2020-00092
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1190A000067
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2020
Initial Date FDA Received12/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-