Model Number P1190A000052
Device Problem Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Date 12/10/2020
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the interface board , foot and head motor needed to be replaced.It is necessary for the progressa¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance disconnect the patient pendant and examine the condition of the connector.Then connect or replace the pendant.Press each of the controls to make sure that they engage the correct function and they do not work intermittently.Each movement must be continuous.Troubleshoot in the event of a doubt.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the interface baord , foot and head motor to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the head of the bed was going up by itself.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
harish vishwanathan
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key11012531
MDR Text Key223183514
Report Number3006697241-2020-00093
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1190A000052
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received12/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse