• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ERC TUBING CLAMP / 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ERC TUBING CLAMP / 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-05-65
Device Problem Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Date 09/29/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 erc tubing clamp / 620mm.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device.The clamp was functional but noisily.The device will be returned to the manufacturer site for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 erc tubing clamp / 620mm was stuck during priming and displayed error codes.The user did not remember which error codes had appeared.There was no patient involvement.
 
Manufacturer Narrative
H.10: the device was returned to the manufacturer site and the reported failure was confirmed.Several parts were replaced to solve the issue and the unit was then shipped back to the customer.Based on the investigation performed for similar cases, the most likely root cause of the reported event is assigned to fluid ingress and use conditions (i.E.Extensive exposure to liquid, e.G.During cleaning).
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ERC TUBING CLAMP / 620MM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key11019105
MDR Text Key221825715
Report Number9611109-2020-00685
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-65
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/16/2020
Initial Date FDA Received12/16/2020
Supplement Dates Manufacturer Received12/22/2020
Supplement Dates FDA Received01/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-