• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CERAMAX CERAMIC INSERT 36 X 54; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CERAMAX CERAMIC INSERT 36 X 54; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS Back to Search Results
Model Number 1218-87-654
Device Problems Fracture (1260); Difficult to Remove (1528)
Patient Problems Pain (1994); Insufficient Information (4580)
Event Date 12/02/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is a revision of the corail and pinnacle hip.This was revised due to the corail stem's subsidence, and the ceramic lined was chipping off.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicated that the patient was experiencing pain that led to the revision surgery.Affected side if left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Photographs of the explanted device appears to depict some material chipping on the face of the liner.It is not possible to determine a root cause using photographic images.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a device history record (mre) review, was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERAMAX CERAMIC INSERT 36 X 54
Type of Device
PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11021022
MDR Text Key221820039
Report Number1818910-2020-27090
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295012528
UDI-Public10603295012528
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1218-87-654
Device Catalogue Number121887654
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/16/2020
Supplement Dates Manufacturer Received12/03/2020
02/24/2021
Supplement Dates FDA Received12/21/2020
02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CORAIL2 NON COL HO SIZE 11.; DELTA CER HEAD 12/14 36MM +5.; PINNACLE 100 ACET CUP 54MM.; CORAIL2 NON COL HO SIZE 11; DELTA CER HEAD 12/14 36MM +5; PINNACLE 100 ACET CUP 54MM
Patient Outcome(s) Required Intervention;
-
-