BOSTON SCIENTIFIC CORK LIMITED SPACEOAR VUE SYSTEM; ABSORBABLE PERIRECTAL SPACER Back to Search Results
Model Number SV-2101
Device Problems Obstruction of Flow (2423); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that spaceoar was attempted to be implanted during a spaceoar placement procedure performed on (b)(6) 2020.Reportedly, the procedure was done under general anesthesia.According to the complainant, during the procedure, it appeared that one side of the y-connector was clogged because the physician was able to pushed the saline through the other side.A second kit was opened however, the y-connector also clogged and the procedure was cancelled.There were no patient complications reported as a result of this event.The patient received planned radiation therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEOAR VUE SYSTEM
Type of Device
ABSORBABLE PERIRECTAL SPACER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORK LIMITED
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key11023012
MDR Text Key226650551
Report Number3005099803-2020-05914
Device Sequence Number1
Product Code OVB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182971
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2021
Device Model NumberSV-2101
Device Catalogue NumberSV-2101
Device Lot Number0010161901
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2020
Initial Date FDA Received12/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-