ROCHE DIAGNOSTICS ELECSYS FREE PSA IMMUNOASSAY; TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA Back to Search Results
Catalog Number ASKU
Device Problem Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/12/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The investigation is ongoing.
 
Event Description
The initial reporter questioned results for 47 patient samples tested for elecsys total psa (total psa) on a cobas 8000 e 602 module.Of the data provided, the results for 23 patient samples were discrepant for total psa.Discrepant results for 1 patient sample tested for elecsys free psa (free psa) was also identified.This medwatch will cover free psa.Refer to medwatch with patient identifier (b)(6)for information on the total psa results.Refer to the attached data for the patient results.It is unknown if the patients were prostatectomy or radiotherapy patients.The questionable results were reported outside of the laboratory.The e602 module serial number was (b)(4).
 
Manufacturer Narrative
No sample material was available for investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FREE PSA IMMUNOASSAY
Type of Device
TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key11023763
MDR Text Key244561158
Report Number1823260-2020-03261
Device Sequence Number1
Product Code MTG
Combination Product (y/n)N
PMA/PMN Number
ASKU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received12/16/2020
Supplement Dates Manufacturer Received11/17/2020
Supplement Dates FDA Received01/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-