Model Number 7205517
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that the arrow on the acl handle has worn off and it is no longer legible; therefore, it is impossible to set the angle of the aimer accurately.The malfunction was found during set up and was solved with no delays or patient harm using a back up device.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.No product batch/lot identification information was provided, thus a manufacturing record review could not be conducted.A complaint history review concluded this was an isolated event.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HDL ASSY DRCTR ACL GDE.
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11026659
MDR Text Key222039357
Report Number1219602-2020-02065
Device Sequence Number1
Product Code NBH
UDI-Device Identifier03596010373250
UDI-Public03596010373250
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7205517
Device Catalogue Number7205517
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/23/2020
Initial Date FDA Received12/17/2020
Supplement Dates Manufacturer Received01/18/2021
Supplement Dates FDA Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-