• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF TRIAL SHELL HANDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF TRIAL SHELL HANDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71362297
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported after a thr procedure, the sterilization services department (ssd) discovered that the screw thread of the handle had sheared.The procedure that was performed previously was completed without delay, using the same device, and no patient injury or other complications were reported.
 
Manufacturer Narrative
Updated lot number.
 
Manufacturer Narrative
Results of investigation: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The returned device has damaged threads rendering the device inoperable.The device was manufactured in 2012 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF TRIAL SHELL HANDLE
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11033161
MDR Text Key222269067
Report Number1020279-2020-07571
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010456540
UDI-Public03596010456540
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71362297
Device Catalogue Number71362297
Device Lot Number12CM13993
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2021
Initial Date Manufacturer Received 11/26/2020
Initial Date FDA Received12/17/2020
Supplement Dates Manufacturer Received11/26/2020
02/04/2021
Supplement Dates FDA Received01/28/2021
02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-