Model Number URF-V2
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2020
Event Type  malfunction  
Manufacturer Narrative
The device have not been returned to olympus medical systems corp.(omsc) but were returned to ocsm for evaluation.The evaluation of the device by ocsm confirmed the following; the bending section rubber was damaged and there was a hole.There was leaked at the bending section rubber.The video connector was corroded by water leak.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus inspected the device at the service department of olympus (b)(4) (ocsm) and found that the angulation control lever of the device did not move and the angulation lock function was not released due to deterioration of the angulation mechanism by rust.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The exact cause of the reported event could not be conclusively determined.It was guessed that the cause of the similar complaint that the angulation was locked and could not be released was that the drum unit in the control section was corroded due to flooding from the place where the leak occurred.Omsc concluded that the reported event may have been caused by corrosion of the drum unit in the control section due to flooding from the leaked bending section rubber, based on similar complaint information and the device inspection result by olympus china sales & marketing.The leakage at the bending section rubber was caused by a broken bending tube.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11034027
MDR Text Key225929072
Report Number8010047-2020-10636
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberURF-V2
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/24/2020
Initial Date FDA Received12/18/2020
Supplement Dates Manufacturer Received01/13/2021
Supplement Dates FDA Received01/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-