• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115309
Device Problems Fluid/Blood Leak (1250); Material Perforation (2205)
Patient Problem No Patient Involvement (2645)
Event Date 11/24/2020
Event Type  malfunction  
Manufacturer Narrative
Prismaflex st100 set ckt has been temporarily approved for use in the us under emergency use authorization#: (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during priming of a prismaflex st100 set, a fluid leak was observed on the effluent line due to a hole.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information added to d10, h3, h6 and h10 h10: the actual device was received for evaluation.A leak test was performed and a leak was observed near the y connector from the replacement pump segment.The reported condition was verified.The cause is a supplier related issue.The complaint has been forwarded to this supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX ST100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11035656
MDR Text Key223763823
Report Number8010182-2020-00303
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414118389
UDI-Public(01)07332414118389
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number115309
Device Lot Number20G1203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2020
Initial Date Manufacturer Received 11/25/2020
Initial Date FDA Received12/18/2020
Supplement Dates Manufacturer Received01/13/2021
Supplement Dates FDA Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-