• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-67A
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2020
Event Type  malfunction  
Manufacturer Narrative
This mdr is result of a retrospective review of complaints.The sensor was investigated, and customer complaint was confirmed via in-vitro qc test as well as after a review of in-vivo data in dms.The in-vitro qc confirmed an intermittent asic crash coupled with noise in both the sig and ref on channels.The investigation shows that the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.
 
Event Description
On 09th august 2020, senseonics was made aware of an incident where user complains of sensor check alerts resulting in an early sensor removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key11041844
MDR Text Key232517900
Report Number3009862700-2020-00599
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491022349
UDI-Public00817491022349
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/20/2020
Device Model Number102096-67A
Device Catalogue NumberFG-4200-00-301
Device Lot NumberWP07468
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Manufacturer Received 08/09/2020
Initial Date FDA Received12/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
-
-