• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315160
Device Problem High Readings (2459)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
The patient's test strips were tested using a retention meter, and controls.Testing results (b)(6).The obtained qc values were in the allowable range routine retention testing is performed.Retention testing data is reviewed, and appropriate actions are taken as needed.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The initial reporter received a questionable result from coaguchek xs meter serial number (b)(4) compare to a laboratory result using dade innovin reagent.The meter result was 2.4 inr at 10:06 am.The laboratory result was 1.86 inr at 10:11 am.The result of 1.86 inr was believed.The therapeutic range is 2.0-3.0 inr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key11045087
MDR Text Key223193645
Report Number1823260-2020-03294
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Catalogue Number04625315160
Device Lot Number44949811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2020
Initial Date FDA Received12/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age69 YR
Patient Weight82
-
-