|
Model Number SYMPHONY DR 2550 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Information (3190)
|
Event Date 12/03/2020 |
Event Type
Injury
|
Event Description
|
The pacing system was implanted on (b)(6) 2006.Reportedly, during a follow-up performed on (b)(6) 2019, the battery impedance was 4.25 kohms and the estimated residual longevity was 33 months.The atrial output was 2.0 v / 0.35 ms for a threshold of 1.0 v / 0.35 ms, the atrial sensitivity was programmed at 0.8 mv and the atrial sensing was measured at 0.8 mv.The atrial lead impedance was measured at 473 ohms.The ventricular output was 2.5v / 0.35 ms in auto for a threshold of 0.75 v / 0.35 ms, the ventricular sensitivity was programmed at 2.7 mv.The ventricular lead impedance was measured at 488 ohms.The next follow-up was scheduled for one year and a half later.During the follow-up performed on (b)(6) 2020, the subject pacemaker was found to have reached its recommended replacement time (rrt) with a battery impedance of 34.83 kohms.The pacemaker was pacing at 70 min-1, with an output of 5.0 v / 0.5 ms.On (b)(6) 2020, the pacing mode changed to ddd during the interrogation.The pacemaker was explanted and should be returned for expertise.Preliminary analysis revealed that, based on the provided patient files, normal battery depletion occurred (according to hrs guidelines).
|
|
Event Description
|
The pacing system was implanted on (b)(6) 2006.Reportedly, during a follow-up performed on (b)(6) 2019, the battery impedance was 4.25 kohms and the estimated residual longevity was 33 months.The atrial output was 2.0 v / 0.35 ms for a threshold of 1.0 v / 0.35 ms, the atrial sensitivity was programmed at 0.8 mv and the atrial sensing was measured at 0.8 mv.The atrial lead impedance was measured at 473 ohms.The ventricular output was 2.5v / 0.35 ms in auto for a threshold of 0.75 v / 0.35 ms, the ventricular sensitivity was programmed at 2.7 mv.The ventricular lead impedance was measured at 488 ohms.The next follow-up was scheduled for one year and a half later.During the follow-up performed on (b)(6) 2020, the subject pacemaker was found to have reached its recommended replacement time (rrt) with a battery impedance of 34.83 kohms.The pacemaker was pacing at 70 min-1, with an output of 5.0 v / 0.5 ms.On (b)(6) 2020, the pacing mode changed to ddd during the interrogation.The pacemaker was explanted and should be returned for expertise.Preliminary analysis revealed that, based on the provided patient files, normal battery depletion occurred (according to hrs guidelines).
|
|
Manufacturer Narrative
|
Please refer to the attached analysis report.
|
|
Search Alerts/Recalls
|
|
|