Generator analysis was performed.The generator was received pulse disabled and in a vboost condition, due to the pulse generator remaining ¿on¿ post explant.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The generator battery measured 1.931 volts and found to be at an eos=yes condition at the time of explant (b)(6) 2020.Burn marks were observed on the pulse generator case, which indicated that the pulse generator may have been exposed to an electro-cautery tool.A visual assessment on the pcba, performed at the pa test bench, revealed no visual anomalies.The battery was remeasured at the completed of the final electrical test as 2.144 v confirming neos=yes.With the pulse generator case removed and the battery still attached to the pcba, the battery measured 2.255 volts.An adjusted blc was performed for the generator and it was concluded that no anomalies exist and the neos condition is an expected event.The reported premature battery depletion was not confirmed.No additional relevant information has been received to date.
|