Model Number X SERIES |
Device Problems
Pacing Problem (1439); Reset Problem (3019)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to pace a patient (age & gender unknown) the device was unable to pace, unable to adjust pacer rate and restarted/rebooted itself multiple times.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The customer's report of "unable to pace/adjust pacer current" was observed during review of the device activity logs.However, the device was put through extensive testing including bench handling and pacer functional stress testing without duplicating the report.The pace/defib engine (processor/bridge/pace and ecg boards) was replaced as a precaution.The customer's report of "reset itself" was not replicated or confirmed.Review of the device activity logs found no evidence of a reset or inappropriate power off event.This report was unsubstantiated.The electrode pads used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
|
|