ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problems Pacing Problem (1439); Reset Problem (3019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to pace a patient (age & gender unknown) the device was unable to pace, unable to adjust pacer rate and restarted/rebooted itself multiple times.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The customer's report of "unable to pace/adjust pacer current" was observed during review of the device activity logs.However, the device was put through extensive testing including bench handling and pacer functional stress testing without duplicating the report.The pace/defib engine (processor/bridge/pace and ecg boards) was replaced as a precaution.The customer's report of "reset itself" was not replicated or confirmed.Review of the device activity logs found no evidence of a reset or inappropriate power off event.This report was unsubstantiated.The electrode pads used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key11049479
MDR Text Key223043727
Report Number1220908-2020-04111
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946019273
UDI-Public00847946019273
Combination Product (y/n)N
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received12/01/2020
Supplement Dates FDA Received02/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-