Model Number 12TLW804F
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/09/2020
Event Type  malfunction  
Manufacturer Narrative
One 12tlw804f catheter was returned for examination.The reported event of inflation difficulty was confirmed.The balloon was found to be ruptured.After relaxing, the latex ruptured edges did not appear to match up.Both the balloon windings were intact.No other visible damage was observed from the catheter.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Balloon rupture and catheter separation, as a result of excessive pull force applied to remove adherent material, are the most frequent causes of reported failures.To minimize the risk of vessel damage, balloon rupture, or tip detachment, the maximum recommended inflation volume and pull force for the catheter should not be exceeded.The arterial embolectomy catheter is not recommended for the removal of fibrous, adherent, or calcified material (e.G.Chronic clot, atherosclerotic plaque).This catheter is not designed to withstand the additional pull force needed to remove these materials.The instructions for use of the product contains the following statement: ¿as with all catheterization procedures, complications may occur.These may include local or systemic infection, local hematomas, intimal disruption, arterial dissection, perforation and vessel rupture, hemorrhage, arterial thrombosis, distal embolization of blood clots and atherosclerotic plaque, air embolus, aneurysm, arterial spasm, arteriovenous fistula formation, and balloon rupture with fragmentation, tip separation and distal embolization.¿ in this case, there was no allegation of patient injury.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that during use in a patient of this fogarty catheter, the balloon could not be inflated.There was no allegation of patient injury.Patient demographics were unable to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY THRU-LUMEN EMBOLECTOMY CATHET
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500294
MDR Report Key11049848
MDR Text Key224441708
Report Number2015691-2020-15156
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K892410
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/05/2021
Device Model Number12TLW804F
Device Catalogue Number12TLW804F
Device Lot Number62290710
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received12/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-