Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 21dec2020.
 
Event Description
It was reported that the ventilator was showing a low priority alarm.There was no patient involvement.The authorized service provider (asp) inspected the device and found an error code indicating blower temperature high.The asp checked that the filters are clean and the fan runs.The asp replaced the blower to address the reported issue, and the unit was returned to use.
 
Manufacturer Narrative
G4:31mar2021.B4:07apr2021.The blower motor assembly was returned for analysis.A visual inspection of the returned component was performed, no notable conditions were found.An investigation was performed and the product analysis technician reported that no fault was found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11050948
MDR Text Key223719346
Report Number2031642-2020-04662
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/24/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received11/24/2020
Supplement Dates FDA Received04/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-