• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
A follow up medwatch will be submitted when additional information becomes available.
 
Event Description
During the pre-surgery inspection, the medical staff found that the double-head pump of the extracorporeal circulation machine reported an error message in cpl mode: stop--.Complaint #(b)(4).
 
Event Description
Complaint (b)(4).
 
Manufacturer Narrative
During the pre-surgery inspection, the medical staff found that the double-head pump of the extracorporeal circulation machine reported an error message in cpl mode: stop.The field service technician (fst) was sent for further investigation.According to service order dated on 2020-12-23 it was found that a data transmission pin was bent on the odu connector at the base of the double-head pump.Re-fixing the data transmission pin, the failure and alarm information are eliminated, and the test operation returns to normal.The device is working properly.Thus the reported failure could be confirmed.The most probable root cause of the reported failure could be determined as a bent pin on the odu connector.The review of the non-conformities during the period of 2015-06-01 to 2020-12-17 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences can be excluded.The reported failure did not happen during patient treatment.The hl20 in question was responsible for this complaint/event.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key11051760
MDR Text Key227111678
Report Number8010762-2020-00446
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701043262
Initial Date Manufacturer Received 12/14/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received12/23/2020
Supplement Dates FDA Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-