• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Electrical /Electronic Property Problem (1198); No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2020
Event Type  malfunction  
Manufacturer Narrative
Since the subject device has been not returned to omsc for the evaluation, the exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that the subject device could not insufflate co2 occasionally at the unspecified timing.At the incoming inspection for the repair, olympus (b)(4) checked the subject device but could not duplicate the reported phenomenon.However it was found that the indicator on the front panel of the subject device was disappeared.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The device was not returned to omsc for evaluation.Omsc reviewed the manufacture history (dhr) of the device and confirmed no irregularity.The exact cause of no insufflation and no image on the front panel could not be conclusively determined.However, the reported phenomena could have been attributed to an abnormality of the internal circuit due to faulty pressure sensor mounted on the main board.The cause of faulty pressure sensor could occurred due to any of the following process errors.Oxygen entered the device and the electrode of the pressure sensor was oxidized and corroded.The wiring inside the pressure sensor was peeled off due to the oxidation corrosion.Foreign matter (epoxy) was adhered due to a mounting mistake of the parts.Because of the adherence of the foreign matter (epoxy), the wiring inside the pressure sensor was peeled off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11052371
MDR Text Key225929959
Report Number8010047-2020-10778
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUHI-4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received02/01/2021
Supplement Dates FDA Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-