Model Number TJF-Q180V |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to omsc but was returned to olympus (b)(4) (ofr).Ofr sent the device to a third party laboratory for microbiological testing.As a result of the testing, the sample collected from all channels of the device tested positive for micrococcaceae (1ufc/endoscope) and the sample collected from distal endo of the device tested positive for gram positive bacteria (2ufc/endoscope).The testing result cleared clear the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the device.[first time] stenothrophomonas maltophilia [second time] pseudomonas aeruginosa and stenothrophomonas malthophilia the device had been reprocessed with a non-olympus automated endoscope reprocessor, wassenburg, using peracetic acid.There was no report of infection associated with this report.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device in this report was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.
|
|