Model Number TJF-Q180V
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc but was returned to olympus (b)(4) (ofr).Ofr sent the device to a third party laboratory for microbiological testing.As a result of the testing, the sample collected from all channels of the device tested positive for micrococcaceae (1ufc/endoscope) and the sample collected from distal endo of the device tested positive for gram positive bacteria (2ufc/endoscope).The testing result cleared clear the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the device.[first time] stenothrophomonas maltophilia [second time] pseudomonas aeruginosa and stenothrophomonas malthophilia the device had been reprocessed with a non-olympus automated endoscope reprocessor, wassenburg, using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device in this report was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11052924
MDR Text Key225930014
Report Number8010047-2020-10793
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170229503
UDI-Public04953170229503
Combination Product (y/n)N
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/26/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received01/14/2021
Supplement Dates FDA Received01/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-