• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71368569
Device Problem Solder Joint Fracture (2324)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  Injury  
Event Description
It was reported that, during a thr surgery, the r3 offset impactor was cracking at the welded junction by the orange handle.No pieces were left inside the patient.Procedure was finished with a smith-nephew back up device.Surgery was not delayed.
 
Manufacturer Narrative
The device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is broken, damaged and has signs of wear and tear from use.The device was manufactured in 2009.According to clinical/medical investigation, this case reports the black impactor tip on r3 offset impactor broke during use.Per complaint details, there was nothing left inside the patient, and all pieces were accounted for.There was no patient harm reported, and the procedure was completed without delay, using a backup device.No further clinical assessment is warranted.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 OFFSET IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11055517
MDR Text Key223138931
Report Number1020279-2020-07680
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010600240
UDI-Public03596010600240
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71368569
Device Catalogue Number71368569
Device Lot Number09CYW00005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Initial Date Manufacturer Received 11/30/2020
Initial Date FDA Received12/22/2020
Supplement Dates Manufacturer Received02/08/2021
Supplement Dates FDA Received02/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-