Model Number 9736242
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system used during a functional endoscopic sinus surgery (fess) procedure.It was reported that the site was having issues verifying their straight suction while using the flat emitter.The site switched to the side emitter and still had some issue with the longer instrumentation.The site needed to move the side emitter very close for patient tracking.The surgeon decided to switch to a different system.After the case it was confirmed that the site had left the flat emitter under the patient when they moved to the side emitter.Even with the flat emitter unplugged, this could cause some interference on the system.Testing the bed also confirmed that there needed to be additional padding underneath the flat emitter to lower the metal interference values.When testing, it was not easy to verify the straight suction when the demo head was resting on the large donut.Technical services recommended off setting the non-invasive patient tracker (nipt) to avoid hitting the electromagnetic (em) threshold with longer instrumentation and this worked.The probable cause was due to metal interference/the operating room set-up.There was a delay of less than one hour and no impact to the patient.Additional information was received.It was reported that the metal interference issue was affecting both verification and tracking.It was noted that it was not communicated to the field that when switching from a flat emitter to a side emitter that it is necessary to remove the unplugged flat emitter from under the patient.The unplugged flat emitter is essentially a slab of metal when the side emitter is plugged in.The tracking issue was able to be resolved by switching the operating room bed that the site uses and the flat and side emitter were functioning properly.This was done after the procedure.The site brought in another system to finish the case as the tracking issue was not resolved and the representative then tested the original system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION FLEXENT
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key11057774
MDR Text Key226033520
Report Number1723170-2020-03379
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9736242
Device Catalogue Number9736242
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received12/22/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
Patient Weight67
-
-