• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Leak/Splash (1354)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's subsidiary, the end-user continued to use the hose by sealing it with plastic tape.Damage was found to the tubing around the hole.
 
Event Description
It was reported that during use of the device for a cardiopulmonary bypass procedure, there was air leaking from the yellow compression gas hose.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Per terumo subsidiary, the hose was replaced at the customer site.Therefore no hose replacement is required.
 
Event Description
Per clinical review: during a cardiopulmonary bypass (cpb) procedure on (b)(6) 2020.It was noticed that there was a tiny leak on the yellow (medical grade gas) line during the procedure.The team used heavy tape and taped up the little crack in the line.The team was able to continue the procedure without issue, without a loss in the fraction of inspired oxygen (fio2) level.There was no delay in the continuation of the surgical procedure.There was no harm or blood loss due to the incident.
 
Manufacturer Narrative
The reported complaint was confirmed via information obtained from the manufacturer's subsidiary.No parts were returned so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11059686
MDR Text Key223778335
Report Number1828100-2020-00499
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/08/2020
Initial Date FDA Received12/22/2020
Supplement Dates Manufacturer Received01/06/2021
02/08/2021
03/02/2021
Supplement Dates FDA Received01/27/2021
03/01/2021
03/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-