• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SMART CABLE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SMART CABLE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0800-0543
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
The customer declined the option to have their glidescope spectrum smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient procedure, using a glidescope spectrum smart cable, the image would disappear intermittently.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE SPECTRUM SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key11060802
MDR Text Key224637965
Report Number9615393-2020-00265
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0800-0543
Device Catalogue Number0800-0544
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2020
Initial Date FDA Received12/22/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-