• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE VIDEO MONITOR (GVM); LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE VIDEO MONITOR (GVM); LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0338
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
The glidescope video monitor (gvm) was returned to verathon for evaluation.A verathon technical service representative evaluated the returned monitor and confirmed the reported image issue.When the verathon technical service representative opened the monitor, the lcd wire harness was found partially disconnected from the main pcba and was not insulated with kapton tape.The verathon technical service representative reconnected the lcd wire harness and insulated it with kapton tape to resolve the issue.Upon completion of the evaluation, the glidescope video monitor (gvm) was returned to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient procedure, using a glidescope video monitor (gvm), there was no image on the display.The customer was able to isolate the issue to the monitor by testing the device with several cables.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE VIDEO MONITOR (GVM)
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key11060974
MDR Text Key224637963
Report Number9615393-2020-00266
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0338
Device Catalogue Number0570-0338
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2020
Initial Date FDA Received12/22/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-