The glidescope video monitor (gvm) was returned to verathon for evaluation.A verathon technical service representative evaluated the returned monitor and confirmed the reported image issue.When the verathon technical service representative opened the monitor, the lcd wire harness was found partially disconnected from the main pcba and was not insulated with kapton tape.The verathon technical service representative reconnected the lcd wire harness and insulated it with kapton tape to resolve the issue.Upon completion of the evaluation, the glidescope video monitor (gvm) was returned to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
The customer reported that during a patient procedure, using a glidescope video monitor (gvm), there was no image on the display.The customer was able to isolate the issue to the monitor by testing the device with several cables.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|